DESCRIPTION:
Cold, rain, snow, dust and humidity associated with use in the field can cause condensation which may damage a meter, rapidly deteriorating its performance and life. HI 9143 is designed for outdoor applications, especially waste water treatment and fish farming. Simple on-site calibration requires no chemical solutions. Just expose the probe to air and press the CAL button. In a few minutes the meter is calibrated and ready to use. Auto-calibration eliminates the need to use a screwdriver or other devices to adjust the reading. The HI 9143, provides readings of O2 (% or mg/l) and temperature and it also supplies the user with compensation of salinity and altitude factors. The meter compensates for temperature effects. HI 9143 is supplied complete with HI 76407/4 D.O. probe with 4 m (13') cable, 2 spare membranes, HI 7041S electrolyte solution, 4 x 1.5V AA batteries and instruction manual in a rugged carrying case.
Dissolved Oxygen Measurement Made Simple!
SPECIFICATIONS:
| SPECIFICATIONS | HI 9143M |
| Range |
mg/l Oxygen 0.00 to 45.00 mg/L % Saturation Oxygen 0.0 to 300.0% Temperature 0.0 to 50.0 °C |
| Resolution |
mg/l Oxygen 0.01 mg/L % Saturation Oxygen 0.1% Temperature 0.1°C |
| Accuracy (@ 20°C/68°F) |
mg/l Oxygen ±1.5% Full Scale % Saturation Oxygen ±1.5% Full Scale Temperature ±0.5°C |
| Calibration | Automatic in saturated air |
| Temperature Compensation | Automatic from 32 to 122°F (0 to 50°C) |
| Auto Altitude Compensation | 0 to 1900m (0 to 6230'), Resolution: 100 meter |
| Auto Salinity Compensation | 0 to 40 g/L (ppt), Resolution: 1 g/L (ppt) |
| Probe | HI 76407/4 Polarographic D.O. probe |
| Battery Type/Life |  : 4 X 1.5 Volt AA / Power plug for 12VDC supply  : approx. 200 hours of continuous use |
| Environment | 32 to 122°F (0 to 50°C) |
| Dimensions | 7.7 x 3.1 x 2.4" (196 X 80 X 60mm) |
| Weight | 3.1 lb (1.4 Kg) |
HOW TO ORDER:
HI9143:Meter supplied with HI76407/4, 4 m (13') cable, 2 spare membranes, HI7041S electrolyte solution, batteries with a carrying case.....$800.00