Automated Aquarium Systems

Amphel pH Electrode


Amphel Probe



DESCRIPTION:

The HI2911B/5 is Hanna's answer to noise related pH measurement problems and the technical inability to use conventional meters with standard pH probes. Now with the AmpHel electrodes you can drive a standard mV meter over longer distance cables, up to 50 meters away without any impedance or noise problems!

Conventional pH Technology:
Due to the high resistance of the glass membrane of a pH electrode, conventional pH measuring systems utilize high impedance signal transmission. Bad insulation of the electrode connectors and cables results in high susceptibility to leakage, stray noise and humidity which tends to give erroneous pH readings. As a result, particular care has to be taken in connecting the electrode to the metering system. For this conventional system, the cable length is restricted to typically less than 10 meters because of the low signal transmission. For measurements to be accurate the use of a high impedance meter is required and it is necessary to provide for high insulation in the meter connections. For these reasons the conventional pH measuring system is delicate.

Breakthrough by HANNA:
With an amplifier built into the electrode, the problems associated with high impedance is now isolated to one location. The high impedance circuitry is now located at the top of the electrode and is completely encapsulated. As a result you now have low output impedance signals from the electrode to the metering system. This means you can use ordinary connectors with long unshielded cables and an ordinary meter. This breakthrough in pH technology provides you with a rugged system for all industrial pH measurements and monitoring.

Characteristics & Adavantages:


SPECIFICATIONS:

  SPECIFICATION   HI2911B/5
  REFERENCE SYSTEM:
  Junction
  Type
  Electrolyte

   Double
   Teflon®
   Polymer
  TEMPERATURE   -5 to 80°C (23 to 176°F)
  MAX PRESSURE   3 bar (43.5 psi)
  LEAD:
  Connector
  Cable

   BNC
   5m (16.5')
  DIMENSIONS   148.5L x 12W mm (5.8 x 0.5")


HOW TO ORDER:

HI2911B/5: AmpHel pH Electrode with preamplified low impedance signal..... $135.00

ACCESSORIES:


HI7004L: pH 4.01 Calibration Solution, 460 mL Bottle.....$Check Pricing
HI7007L: pH 7.01 Calibration Solution, 460 mL Bottle..... $Check Pricing
HI7010L: pH 10.01 Calibration Solution, 460 mL Bottle....$Check Pricing
HI70300L: Electrode Storage Solution, 460 mL Bottle...... $Check Pricing





How To Place an Order by Phone, eMail, or Fax

VISA/MC Accepted




[ HOME | PRODUCTS | HANNA | MILWAUKEE | LAMOTTE | MONITORING & CONTROL SYSTEMS | TECHNICAL INFO | SALES | SERVICE | CONTACT US ]